29
c)
Beckman pH meter, Model 340
i)
Press CAL
to initiate the calibration mode.
ii)
Press CLEAR
twice to clear the previous calibration.
iii)
Wash the electrode thoroughly with DIH2O
and immerse in pH 7.00 standard buffer
solution. Push READ
and swirl the solution around the bulb of the electrode to speed
equilibration.
This calibration point has been accepted
when the eye stops flashing.
iv)
Again, wash the electrode
thoroughly with DIH2O
and immerse in pH 4.00 standard
buffer solution.
Push READ and swirl the solution
around the bulb of the electrode to
speed equilibration.
This calibration point has been accepted
when the eye stops flashing.
Record the slope displayed.
v)
Press CAL
to end the calibration procedure.
vi)
Rinse and dry the electrode and store
it in the storage buffer until required.